Model Number W1DR01
Device Problem False Alarm (1013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/17/2023
Event Type  malfunction  
Manufacturer Narrative
Continuation of d10: model: 5076-52 lead; implanted:(b)(6) 2003.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that the right atrial (ra) lead exhibited blanking every other atrial flutter wave, causing right ventricular (rv) lead pacing at higher rates due to tracking.The right atrial (ra) lead also exhibited a low lead impedance warning.Both leads remain in use. no patient complications have been reported as a result of this event.It was further reported that the low impedance on the ra lead was due to a false low impedance warning by the implantable pulse generator (ipg).The ipg remains in use.
 
Manufacturer Narrative
Corrected h6 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a false alert for lead impedance out of range.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURE XT DR MRI SURESCAN
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN   486056
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19190890
MDR Text Key341132856
Report Number3008973940-2024-03052
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberW1DR01
Device Catalogue NumberW1DR01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/11/2024
Initial Date FDA Received04/26/2024
Supplement Dates Manufacturer Received04/29/2024
05/02/2024
Supplement Dates FDA Received05/02/2024
05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4965-35 LEAD
Patient Age80 YR
Patient SexFemale
-
-