This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: g2: (health professional should be unmarked).The device was returned to olympus for inspection, and based on the results of the investigation.The customer's allegation was confirmed.From investigation result, it was estimated, that an error was detected in the pressure sensor on the main board.And the air feeding operation was stopped, while the alarm sounds and led other than the power supply was turned off.Additionally, the secondary line leak was found to have failed test.And it was presumed, that leakage occurred from the conduit, due to a failure of the electro-pneumatic proportional valve serving as the secondary conduit or the manifold u, leading to this event.A definitive root cause could not be identified.Olympus will continue to monitor field performance for this device.
|