• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION GZ-130PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION GZ-130PA; TRANSMITTER Back to Search Results
Model Number GZ-130PA
Device Problem Patient Data Problem (3197)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Event Description
The biomedical engineer (bme) reported that the ecg waves are intermittently dropping from the gz transmitter and at the remote nurse's station (rns) so the biomed swapped it out with another gz transmitter on the patient, no patient harm was reported.Biomed reported they ran the device on a simulator and it still continued to drop ecg waves.No physical damage to the unit reported.Biomed requested a warranty exchange unit and will return this gz transmitter to nihon kohden.
 
Manufacturer Narrative
The biomedical engineer (bme) reported that the ecg waves are intermittently dropping from the gz transmitter and at the remote nurse's station (rns) so the biomed swapped it out with another gz transmitter on the patient, no patient harm was reported.Biomed reported they ran the device on a simulator and it still continued to drop ecg waves.No physical damage to the unit reported.Biomed requested a warranty exchange unit and will return this gz transmitter to nihon kohden.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GZ-130PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 116-8 560
JA  116-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key19192831
MDR Text Key341493042
Report Number8030229-2024-04235
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921117415
UDI-Public04931921117415
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153707
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGZ-130PA
Device Catalogue NumberGZ-130PA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/26/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
RNS
-
-