Catalog Number 2C4063K
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2024
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter first name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume folfusor leaked from an unspecified location.This was noticed while unpacking the pump from the shipping box.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: d4 expiration date, d9, h3, h4, h6, h11.H4: the lot was manufactured august 1, 2023 to august 3, 2023.H11: the actual device was received for evaluation.Visual inspection observed fluid inside the bag that contained the sample suggesting a leak may have occurred.Further inspection confirmed that the leak was caused by a damaged blue winged luer cap.Indentation marks from the manufacturing flow wrap sealer equipment were observed on the damaged area of the blue winged luer cap.The reported condition was verified.The cause of the damaged cap was determined to be related to the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19193549
MDR Text Key341308584
Report Number1416980-2024-01918
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080161
UDI-Public(01)00085412080161
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4063K
Device Lot Number23H001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/26/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/30/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-