Additional information: d4 expiration date, d9, h3, h4, h6, h11.H4: the lot was manufactured august 1, 2023 to august 3, 2023.H11: the actual device was received for evaluation.Visual inspection observed fluid inside the bag that contained the sample suggesting a leak may have occurred.Further inspection confirmed that the leak was caused by a damaged blue winged luer cap.Indentation marks from the manufacturing flow wrap sealer equipment were observed on the damaged area of the blue winged luer cap.The reported condition was verified.The cause of the damaged cap was determined to be related to the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|