Model Number MAJ-2111
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported the connecting tube had a broken attachment clip and unable to connect to reprocess (oer).There was no damage to the reprocessor just the plastic connection piece.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.New information added to the following fields: d8, d9, h3, h6, h11.Corrected fields: d4: lot number leave on blank.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The device was returned to olympus for inspection, and the customer's complaint was not confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that "connecting tube could not be connected to the connector at reprocessing basin due to breakage of the tube" occurred due to external stress was applied to lock lever of the connecting tube, which led to breakage of the tube.The event can be detected by following the instructions for use which state: preparation and inspection visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.Move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.Olympus will continue to monitor field performance for this device.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The dhr was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The case is not due to design.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to external stress applied to the lock lever of the connecting tube, which led to breakage of the tube.Subsequently, the tube was unable to be connected.The user may have applied stress toward wrong direction which caused the external stress.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Type of Device
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19195310
MDR Text Key341251955
Report Number9610595-2024-08767
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170404061
UDI-Public04953170404061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190969
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-2111
Device Lot Number41A
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/26/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/31/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-