SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2024
Event Type  malfunction  
Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the xenon light source lamp did not light up.The bulb exploded within the scope, causing the glass behind the bulb to break.The issue occurred during an unspecified procedure.There were no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the manufacturer¿s investigation.A device evaluation, a review of the device history record (dhr), a review of the instructions for use (ifu), as well as a historical trending analysis were conducted during this investigation.The subject device was returned, and the user¿s report was confirmed.The bulb had exploded, the turret lens was broken, and the scope socket slider was worn out ¿ causing the failure of high intensity mode.Additionally, there was a non-olympus bulb installed on the subject device.The review of the dhr did not find any abnormalities or anomalies identified during production.The device met all specifications upon release.The ifu contains the following information related to the reported failure: ¿do not apply excessive force to the light source and/or other instruments connected.Otherwise, damage and/or malfunction can occur.¿ ¿never install a lamp that has not been approved by olympus.The use of a nonapproved lamp can cause damage to the light source and ancillary equipment, malfunction, or a fire.¿ based on the results of the investigation as well as the device evaluation results, it is likely the undue stress on the subject device caused the reported failure mode.However, a definitive root cause could not be determined.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19196678
MDR Text Key341792893
Report Number3002808148-2024-03938
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/19/2024
Initial Date FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-