It was reported that during an atrial fibrillation (a fib) procedure involving an farawave catheter, the physician experienced difficulty moving the guidewire while attempting to cannulate right superior pulmonary vein (rspv).The non-boston scientific guidewire became stuck while catheter tip was floating in the chamber, so it is not believed to be tissue entrapment.After attempting and failing to free the guidewire, the system was removed, and the guidewire and catheter were replaced.To remove the guidewire, it was needed to remove from distal end of catheter and wire was ripped.For imaging fluoro and ice were used, also the case was performed on non-interrupted anticoagulant.Heparin was given pre transeptal and the act results prior to the event were above 300.After the device was replaced, the procedure was completed without patient complications, the catheter was disposed so it will not be returned.
|