• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FARAWAVE PULSED FIELD ABLATION CATHETER; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FARAWAVE PULSED FIELD ABLATION CATHETER; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER Back to Search Results
Lot Number 0033476252
Device Problem Entrapment of Device (1212)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2024
Event Type  malfunction  
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event Description
It was reported that during an atrial fibrillation (a fib) procedure involving an farawave catheter, the physician experienced difficulty moving the guidewire while attempting to cannulate right superior pulmonary vein (rspv).The non-boston scientific guidewire became stuck while catheter tip was floating in the chamber, so it is not believed to be tissue entrapment.After attempting and failing to free the guidewire, the system was removed, and the guidewire and catheter were replaced.To remove the guidewire, it was needed to remove from distal end of catheter and wire was ripped.For imaging fluoro and ice were used, also the case was performed on non-interrupted anticoagulant.Heparin was given pre transeptal and the act results prior to the event were above 300.After the device was replaced, the procedure was completed without patient complications, the catheter was disposed so it will not be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARAWAVE PULSED FIELD ABLATION CATHETER
Type of Device
CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key19197265
MDR Text Key341479062
Report Number2124215-2024-25399
Device Sequence Number1
Product Code QZI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Lot Number0033476252
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/08/2024
Initial Date FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2024
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
-
-