This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b3, d1, d2, and d4 (model and catalog).Device evaluation: zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including functional testing, defib/pacer stress testing, bench handling, and defib cycling, using the returned one-step pacing cable, without duplicating the report.The device was recertified and returned to the customer.Review of the first clinical file observed the device displayed a clear ecg signal on pads before the device was turned to pacer mode.When the device was turned to pacer mode, the device detected an ecg lead-off message.Ecg lead off is an expected prompt if 1 or more ecg p leads are not properly coupled to the patient.The device detected valid impedance for electrode pad contact and the device was able to capture pacing activity until the end of the case.Per the design of the r-series, an ecg signal through leads is required to capture pace.Review of the second clinical file, showed the device was turned to pacer mode, the device went to p3 lead and captured pacing as expected.After 12 seconds, there was noise with the leads, but the pacer function was active and the device continued to capture pacing activity (with noise).After 3 minutes the user changed the lead view to p1 and no noise was seen and clean capture was observed.No trend is associated with reports of this type.
|