Model Number X SERIES |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was evaluated by zoll medical corporation.The customer's report was duplicated and attributed to a faulty transient voltage suppressor diode on the monitor board.The monitor board was replaced to resolve the report.The customer was sent a replacement device.Analysis for reports of this type has not identified an increase in trend.
|
|