• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR Back to Search Results
Model Number M3535A
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2024
Event Type  malfunction  
Event Description
It was reported to philips that the heartstart mrx defibrillator exhibited noise in the power supply.Patient involvement is unknown at this time, however, no adverse patient impact or injury was reported by the customer.Additional information has been requested.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Manufacturer Narrative
This report is based on information provided by philips remote service engineer (rse) and has been investigated by the philips complaint handling team.Philips received a complaint on a heartstart mrx device indicating that the device had a noise in the power supply.The customer was informed that service could no longer be completed on the unit as the device had reached end of life (eol).The heartstart mrx device and all associated service/support was discontinued on 03-feb-2022.As troubleshooting was not completed for the device, the reported issue could not be verified, and a cause could not be established.The reported problem was not confirmed.Based on the information available and results of additional analysis, no further action is necessary at this time.The customer was made aware of the end of life terms and the device remains at the customer site.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tanya deschmidt
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key19201232
MDR Text Key341258333
Report Number3030677-2024-01551
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/10/2024
Initial Date FDA Received04/29/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/20/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-