CARDIAC SCIENCE CORPORATION POWERHEART G5 KIT,AUTO,US EN/FRN CA, CPR; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number G5A-90C |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to a faulty integrated circuit on the main board.The main board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|
Event Description
|
Complainant alleged that during functional testing, the device prompted an "hv cap idle test failure" message.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
|