Model Number ALLURA XPER FD20 OR TABLE
Device Problem Radiation Output Failure (4027)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Event Description
It has been reported to philips that the wireless foot switch would not work.There was no reported patient or user harm.Due to the lack of information, we are conservatively reporting this event.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to additional information received, the system was not in clinical use when the issue was identified.A philips remote service engineer (rse) inspected the system remotely and confirmed that the wireless footswitch was not working.The customer requested for the part id to order and replace the wireless footswitch.No further information regarding the issue has been provided and no onsite service has been performed by philips it was not requested by the customer.The same issue reoccurred.The philips engineer suggested to replace the wfs and base station in the subsequent case.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
derrick massuri
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key19201929
MDR Text Key341286947
Report Number3003768277-2024-02500
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059085
UDI-Public00884838059085
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20 OR TABLE
Device Catalogue Number722023
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/29/2024
Date Device Manufactured08/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-