Livanova deutschland received a report that, during a case, the cardioplegia side of a heater-cooler system 3t was not cooling and perfusion experienced power surge in the or.There is no report of any patient injury.After taking out the cardioplegia bridge noticed the coil was leaking refrigerant into the tank and bubbles were coming out of the coil on the cardioplegia cold tank.Also, when plugging in the 3t and turning it on in the or room would get high current alarm coming from the isolated power panel in the room.
|