BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4711K
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Event Description
It was reported two (2) small volume folfusors disconnected.Prior to preparation of the device a separation of unspecified parts was observed (no further details were provided).There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter facility name: (b)(6).E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information: d9, h3, h4, h6, and h11.H4: the lot was manufactured between august 18, 2023 and august 21, 2023.H11: two devices were received for evaluation.The first sample did not have fluid in its bladder, while the second sample had fluid in its bladder.Visual inspection was performed and observed that coil cap of first sample was separated from the housing; lug diameter of the housing was found to be slightly lower than the specification.The coil cap from the second sample was observed to be securely attached to its housing.The reported condition was verified in one returned sample.The cause of the condition could be the slightly smaller lug diameter.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19202015
MDR Text Key341496981
Report Number1416980-2024-01935
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412082172
UDI-Public(01)00085412082172
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4711K
Device Lot Number23H012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/29/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-