Additional information: d9, h3, h4, h6, and h11.H4: the lot was manufactured between august 18, 2023 and august 21, 2023.H11: two devices were received for evaluation.The first sample did not have fluid in its bladder, while the second sample had fluid in its bladder.Visual inspection was performed and observed that coil cap of first sample was separated from the housing; lug diameter of the housing was found to be slightly lower than the specification.The coil cap from the second sample was observed to be securely attached to its housing.The reported condition was verified in one returned sample.The cause of the condition could be the slightly smaller lug diameter.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|