Catalog Number CS-12122-F |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/20/2024 |
Event Type
malfunction
|
Event Description
|
It was reported "in icu, the doctor found the swg kinked during use on the patient." no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
(b)(4) the report of a kinked guide wire could not be confirmed through examination of the returned sample.The customer returned an opened hemodialysis kit including one guide wire within its advancer, an arrow raulerson syringe (ars), and 18ga introducer needle for evaluation.The guide wire was returned within the advancer tube and signs of use were observed on the guide wire and its assembly.No defects or anomalies were observed on the returned guide wire.The guidewire passed the dimensional and functional test specifications.A device history record review did not identify any manufacturing related issues.Based on the customer description and the sample received, no problem was found on the returned sample.Teleflex will continue to monitor and trend for reports of this nature.
|
|
Event Description
|
It was reported "in icu, the doctor found the swg kinked during use on the patient." no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
|
|
Search Alerts/Recalls
|