MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB10013
Device Problems Application Program Problem: Parameter Calculation Error (1449); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2024
Event Type  malfunction  
Event Description
It was reported that during a cori assisted tka surgery, the real intelligence robotic drill could not form a stable connection."critical error" message was displayed when running a drill diagnostic and when starting a new case, it could not proceed past initial drill registration either.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event Description
It was reported that during a cori assisted tka surgery, the real intelligence robotic drill could not form a stable connection."critical error" message was displayed when running a drill diagnostic and when starting a new case, it could not proceed past initial drill registration either.But then in the case there were a few weird glitches, a random spike while doing initial stresses& a rotated femur when setting the axis.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H3, h6: the real intelligence robotic drill, part # rob10013, serial # sn000112, used for treatment, was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was confirmed.It was found that the thermistor wires have been severed internally, causing errors when attempting to set up the drill.A visual inspection of the images/videos provided by the customer was performed.The reported problem was confirmed.A visual inspection of the image was conducted and confirmed a critical error was displayed.The most likely cause of the reported errors seems to have been due to severed thermistor wires in the drill cable, or a faulty thermistor.It is possible that this occurred due to twisting or bending of the cable.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of the reported complaint.The failure mode and associated risk have been anticipated within the risk file including a documented risk level.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19202530
MDR Text Key341291166
Report Number3010266064-2024-00096
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB10013
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/10/2024
Initial Date FDA Received04/29/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-