Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Run on Battery (1466)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2024
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the v60, indicating that the device was displaying a battery failure error.It was reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the authorized service provider (asp) and confirmed the reported problem.It was recommended to replace the battery to resolve the reported issue.Investigation is ongoing.
 
Manufacturer Narrative
E1: reporting institution name:(b)(6).
 
Manufacturer Narrative
Per good faith effort (gfe) response received 03jun2024, no repairs were completed by a manufacturer's product support engineer (pse) as the customer declined service and repair due to the device not being under warranty.No additional details regarding the reported issue and its resolution are available.The investigation concludes that no further action is required at this time.The device remains at the customer site.If the decision is made to have the device evaluated and repaired a new service order will be opened and will be captured through philip's normal complaint procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19202909
MDR Text Key341284033
Report Number2518422-2024-23353
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/20/2024
Initial Date FDA Received04/29/2024
Date Device Manufactured04/13/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-