Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0902 captures the reportable event of gauge reading inaccurate.Block h11: investigation results: the returned alliance ii inflation syringe was analyzed, and a visual inspection found that the gauge needle was above the upper limit which is 12 atm.Functional testing was attempted to be performed but due to the returned device with the needle above the upper limit, the gauge needle could not advance further.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of gauge reading inaccurate was confirmed.The results of the analysis performed on the returned device found that the needle was above the upper limit, which is 12 atm, therefore with an inaccurate reading.It is possible that the way in which the device was handled and manipulated may have caused the encountered problem.Such as a possible hit on the gauge or an accidental drop of the device.Therefore, the most probable root cause is an adverse event related to procedure.
|