• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problems Device Alarm System (1012); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2024
Event Type  malfunction  
Manufacturer Narrative
H10: g: phone: (b)(6).
 
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had a battery issue.There was no patient involvement when the issue occurred.No patient or user harm reported.Investigation is ongoing.
 
Manufacturer Narrative
A follow up was performed with the key market, and it was reported that the device had a 1115 error code (check vent: auxiliary alarm supply failed), and that the device did not have a battery issue.The km responder reported that the customer declined to have the device repaired by philips.Insufficient information is available to determine the resolution of the event.No further details could be obtained regarding the event, and it could not be determined if the customer's issue was resolved or if the device had been repaired.No parts were returned for failure investigation.In the event that parts are returned or if new information is provided, the complaint will be reopened to update the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19203629
MDR Text Key341294080
Report Number2518422-2024-23392
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2024
Initial Date FDA Received04/29/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/29/2024
Date Device Manufactured04/09/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-