• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE; DQX WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Catalog Number HPWA-35-180
Device Problem Flaked (1246)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/17/2024
Event Type  malfunction  
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1: customer name and address = (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
As reported, during an unknown procedure, the coating of a roadrunner uniglide hydrophilic wire guide ¿curled¿.Latex free gloves were worn, and the coating was activated with normal saline.The wire was used four-to-five times.The wire was not kept hydrated when not in use but was reactivated before re-use.The anatomy was not tortuous, scarred, or calcified.Resistance was not encountered upon insertion of the wire guide.Then wire was not removed through a needle or other metal instrument.Reportedly, the coating was uneven and not smooth and curled in the middle of the wire ¿like waves¿.No metal was exposed.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE
Type of Device
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key19207189
MDR Text Key342013371
Report Number1820334-2024-00584
Device Sequence Number1
Product Code DQX
UDI-Device Identifier00827002561748
UDI-Public(01)00827002561748(17)260731(10)CINC004739
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K110009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHPWA-35-180
Device Lot NumberCINC004739
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/17/2024
Initial Date FDA Received04/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-