• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMPREHENSIVE REVERSE SHOULDER CENTRAL SCREW; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMPREHENSIVE REVERSE SHOULDER CENTRAL SCREW; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 115396
Device Problem Migration (4003)
Patient Problem Bruise/Contusion (1754)
Event Date 02/15/2024
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10: concomitant medical products, part number (lot number): ¿ 110027734 (66500759) ¿ 110031378 (66500762) associated product information: ¿ 113633 (64964350) ¿ 180554 (unknown) ¿ 180554 (unknown) ¿ 180555 (unknown) ¿ 180555 (unknown) ¿ 180556 (unknown) ¿ 405800 (66394301) ¿ 110010066 (66500764) ¿ 110030777 (66072611) ¿ 110031400 (66394125) ¿ 110031429 (65434183) this report was initially reported under the incorrect mfr number (0001822565-2024-00901).The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that a few days following an initial arthroscopy procedure, the patient experienced implant dissociation.The patient underwent additional surgery approximately one (1) week following the original procedure, and the surgeon discovered that the central screw wasn't fully tightened, causing the looseness.The screw was adjusted, and other concomitant devices were retained during the revision, with no other patient harm reported.Attempts have been made, and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE REVERSE SHOULDER CENTRAL SCREW
Type of Device
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19207292
MDR Text Key341322196
Report Number0001825034-2024-01095
Device Sequence Number1
Product Code PHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193373
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number115396
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/03/2024
Initial Date FDA Received04/29/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
H11
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
-
-