One device was received for evaluation.Visual inspection found a scratched lcd lens, and fluid ingression on the dso sensor.There was no evidence in the event history log.Functional testing was able to duplicate the issue.It was determined that the extensive fluid ingression was the root cause.The main board, cam flex sensor, led flex sensor, lcd lens, dso sensor and seal, front and rear piezo assembly were all replaced.The service history review had no indication that the complaint was related to a service of the device within the review period.
|