• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had "no ac (alternating current) inlet.There was no patient involvement when the issue occurred.No patient or user harm reported.Investigation ongoing.
 
Manufacturer Narrative
A service engineer (se) evaluated the device and reported that the power supply was defective.The se then replaced the power supply, and the issue was resolved.The system meets the specification for the performed service and is returned to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19209845
MDR Text Key341366924
Report Number2518422-2024-23637
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838009851
UDI-Public00884838009851
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2024
Initial Date FDA Received04/29/2024
Supplement Dates Manufacturer Received04/28/2024
Supplement Dates FDA Received05/03/2024
Date Device Manufactured12/03/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-