• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-ELITE
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2024
Event Type  malfunction  
Manufacturer Narrative
The field service engineer replaced a1 connector and verified it in according to original equipment manufacturers (oem) instructions at the customer facility.The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported the endoscope reprocessed a1 port was broken.There were no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected field: h6 (component code is being corrected to 4733 - connector/coupler additional information added to field d9, h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 years since the subject device was manufactured.An olympus field service engineer (fse) was dispatched to the user facility and the customer's allegation of a1 port broken was confirmed.Based on the results of the investigation, it is likely that the user hit the connector something hard or stress to the connector toward loosening direction was accumulated, leading to breakage of the connector.However, root cause of the suggested event could not be identified.The event can be detected/prevented by following the instructions for use which state: chapter 5 inspection and preparation before use.5.6 inspecting the connectors.Check the following for each connector.¦ the connector should be fixed firmly.¦ the o-rings should be free of irregularities such as cracks, tears, or dents.If any irregularity is found, do not use the reprocessor and contact olympus.Do not use the equipment if any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral devices or facilities near the equipment.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19210060
MDR Text Key341818124
Report Number9610595-2024-08874
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201920
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-ELITE
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/15/2024
Initial Date FDA Received04/29/2024
Supplement Dates Manufacturer Received05/20/2024
Supplement Dates FDA Received05/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-