MAUDE Adverse Event Report: ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Manufacturer Narrative
The reported complaint that the autopulse platform (b)(6) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) was confirmed during functional testing and review of the archive data.The root cause of the (ua) 07 was due to a failure of single point load cell #2, likely attributed to failed component(s), or mishandling such as a drop.Visual inspection of the returned platform was performed, and no physical damage was observed.A review of the archive data showed (ua) 07 messages; thus, confirming the reported complaint.The autopulse platform failed the initial functional testing due to the user advisory (ua) 07 error message displayed upon powering on.The load cell characterization test indicated load cell # 2 was over-reporting.Single point load cell #2 will be replaced to address the observed failure.Upon service completion, the platform will be subjected to final functional testing to ensure that the device passes all testing criteria.Historical complaints were reviewed for service information related to the reported complaint, and there were no similar complaints reported for autopulse platform with (b)(6).
 
Event Description
The customer reported during a training session, the autopulse platform (b)(6) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large).Per the customer, the training occurred a couple days after it was used on a patient where it worked perfectly.In between these two uses it was not used, dropped or damaged in any way.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key19210418
MDR Text Key341434148
Report Number3010617000-2024-00350
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-