MEDTRONIC SINGAPORE OPERATIONS ENSURA DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number EN1DR01 |
Device Problems
Electromagnetic Interference (1194); Low impedance (2285)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/22/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that due to one time miss-measurement of the implantable pulse generator (ipg), the right atrial (ra) lead and right ventricular (rv) lead exhibited low impedance and polarity switch.It was suspected that the electromagnetic interference was the cause.Ipg system remans in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the right ventricular pacing impedance was below the expected lower range.Analysis of the device memory indicated the atrial pacing impedance was below the expected lower range.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|