C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, KIT, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 7707540 |
Device Problem
Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/05/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Event Description
|
It was reported that sometime post a port placement, when the physician removed the port of a patient, there was allegedly a kink.Reportedly, the port was removed percutaneously.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that sometime post a port placement, when the physician removed the port of a patient, there was allegedly a kink.Reportedly, the port was removed percutaneously.There was no reported patient injury.
|
|
Manufacturer Narrative
|
Manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one x-port isp implantable port attached to a groshong catheter was returned for evaluation.Functional, gross visual, tactile and microscopic visual evaluations were performed.Gross examination of the port body showed multiple puncture access of the port septum and distinct curvature of the attached catheter.A kink was noted to the attached catheter.Therefore the investigation is confirmed for the reported deformation issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that sometime post a port placement, when the physician removed the port of a patient, there was allegedly a kink.Reportedly, the port was removed percutaneously.There was no reported patient injury.
|
|