Zoll medical corporation evaluated the device and the customer's report for no power up was not replicated or confirmed.The device was put through extensive testing including bench handling and power cycling using the customer's returned batteries resulted in no findings.The returned batteries were able to power up the device.No power up condition would not be captured in the device log.The reported problem for critical error 5 was observed during review of the device data logs.However, the device was put through extensive testing including bench handling, power cycling, and functional stress testing without duplicating the report.The main board was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|