DEPUY MITEK LLC US ARTHRO CLEVERHOOK-LEFT *EA; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Catalog Number 214641
Device Problems Break (1069); Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2024
Event Type  malfunction  
Event Description
It was reported by the sales rep from switzerland that on 4/18/2024 the arthro cleverhook device was broken, the device jaw was broken and did not work properly anymore.The device was new.There was no reports of delay reported.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.H4: the device manufacture date was unknown udi:(b)(4).
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4: the date of manufacture has been added.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the investigation has been completed, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the product was returned to depuy synthes for evaluation.Depuy synthes then conducted visual inspection of the devices received.The device was received and evaluated.When performing the visual inspection, it could be observed that the device comes in used condition, the distal part of the device was found to be dull and nicked, also it was identified that the jaw was loose.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the arthro cleverhook-left *ea would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life.As per ifu-107066 end of useful instrument life is generally determined by wear or damage from handling or surgical use.Inspect instruments between uses to verify proper functioning.Visually inspect the instrument and check for damage and wear.Jaws and teeth should align properly.Cutting edges should be free of nicks and have a continuous edge.Therefore, it has been determined that no corrective and/or preventative action is required.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19212203
MDR Text Key341465406
Report Number1221934-2024-01406
Device Sequence Number1
Product Code NBH
UDI-Device Identifier10886705004225
UDI-Public10886705004225
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number214641
Device Lot Number23C02
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/21/2024
Initial Date FDA Received04/30/2024
Supplement Dates Manufacturer Received04/30/2024
Not provided
Supplement Dates FDA Received05/14/2024
05/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-