• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 1000 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 1000 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 1000
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2024
Event Type  malfunction  
Manufacturer Narrative
Stryker evaluated the customer's device and was unable to duplicate the reported issue.The device can no longer be repaired.The customer was provided with a replacement device.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event Description
A customer contacted stryker to report that their device would not power on.As a result, defibrillation therapy would not be available, if needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 1000 DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key19212579
MDR Text Key341395136
Report Number0003015876-2024-01362
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873815465
UDI-Public00883873815465
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122600
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000
Device Catalogue Number99425-000025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2024
Initial Date Manufacturer Received 04/05/2024
Initial Date FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-