VERATHON MEDICAL ULC GLIDESCOPE AVL MONITOR; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0304
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2024
Event Type  malfunction  
Manufacturer Narrative
The customer was unable to have their glidescope avl monitor returned to verathon for evaluation due to this device reaching its end-of-life date.Since the device was not returned to verathon the cause could not be determined.Review of complaint history for the video monitor serial number "(b)(6)" did not identify any previous complaints reported to verathon.Trending analysis for the glidescope avl monitors does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
A customer reported that during a patient procedure, using a glidescope avl monitor, the image went out.No delay in the procedure, use of a backup device, or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE AVL MONITOR
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key19213703
MDR Text Key341485994
Report Number9615393-2024-00066
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0304
Device Catalogue Number0570-0314
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2024
Initial Date FDA Received04/30/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/11/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-