Model Number 866389 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/03/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Event Description
|
The customer reported that the system is not producing an audible alarm.The device was in use on patient at time of event, there was no adverse event reported.
|
|
Manufacturer Narrative
|
A philips remote service engineer (rse) spoke with a customer onsite to evaluate the device in question.The rse indicated during an evaluation of the system that the system has no audible alarms due to the system speaker being unseated.The rse directed the customer to re-seat the device speaker to resolve their issue.The philips remote service engineer (rse) confirmed the customers device issue and resolve the device issue by reseating the system speaker.After the system speaker was re-seated the system was placed back into service.
|
|