Model Number V60 V60PLUS VENTILATOR
Device Problem Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that the device could not be powered on.The device was reported to be outside of use at the time of the reported problem.No patient or user harm reported.The customer informed the remote service engineer (rse) that they could not power on the device.The rse provided the customer with the latest service manual revision t.The rse instructed the customer to test for the 24 volts direct current (dc) that should be going to the power management (pm) printed circuit board assembly (pcba).If there was no voltage found/measured, then the rse advised replacing the pm pcba.The customer informed the rse that they disconnected and reconnected the power cord and the battery, and the device powered on fine.The customer tried many times to recreate the initial unable to power on issue but was unable to.The customer stated that they tested the ventilator and it passed successfully.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19214109
MDR Text Key341402278
Report Number2518422-2024-23867
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2024
Initial Date FDA Received04/30/2024
Date Device Manufactured12/22/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-