Philips has investigated this complaint.According to the additional information collected, the system was not in clinical use at the time of the event occurred.Philips field service engineer (fse) inspected the system onsite and confirmed that there was disk bay issue.A review of the system logfiles found disk bay error.Due to manufacturing issues, the disk bay may not function as intended, leading to issues with the system's nehalem pc which is used for host pc, ip-pc and the flexvision pc.If one of these pcs break down, the system stops functioning and no imaging is possible.Philips has initiated a field safety corrective action (2023-igt-bst-027).Upon troubleshooting actions, fse replaced the disk bay.After replacement, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
|