• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Run on Battery (1466)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/23/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that there was a battery failure alarm.The device was reported to be outside of use at the time of the reported problem.No patient or user harm reported.The device was found to be working fine without the battery installed, so the authorized service provider (asp) recommended to the customer to replace the battery.This investigation is ongoing.
 
Manufacturer Narrative
The device was found to be working fine without the battery installed, so the authorized service provider (asp) recommended to the customer to replace the battery.In a good faith effort (gfe) response from the authorized service provider (asp) received on 08may2024, it was stated that the customer had decided to use a 3rd party.No further information regarding the repair or current status of the device was communicated.No further work has been performed by philips.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19214466
MDR Text Key341409070
Report Number2518422-2024-23899
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public(01)00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/23/2024
Initial Date FDA Received04/30/2024
Supplement Dates Manufacturer Received05/08/2024
Supplement Dates FDA Received05/09/2024
Date Device Manufactured04/07/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-