Model Number 97800
Device Problems Energy Output Problem (1431); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Device Overstimulation of Tissue (1991); Electric Shock (2554)
Event Date 03/07/2024
Event Type  malfunction  
Event Description
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for urge incontinence, urinary/bowel dysfunction.It was reported that the therapy is not effective at all since implant date.Patient stated for the first week after the implant it was giving him shocks so he turned it down.Patient said he has been in touch with his healthcare provider and medtronic representative.They mentioned that they turned it up the settings a little bit today but still no results.The patient was redirected to their healthcare provider to further address the issue.Additional information was received from a manufacture representative.The rep reported the patient mentioned the stim felt too strong when i received a whatsapp call from him as he is out of the country.I told him to turn the stim down to a more comfortable setting.The patient turned the stim down a couple notches and said it was comfortable.When the rep was asked for resolution they reported patient has not reported and strong stim feeling or irritating sensations from the stim.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM X
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key19215268
MDR Text Key341468464
Report Number3004209178-2024-10034
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00763000484668
UDI-Public00763000484668
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number97800
Device Catalogue Number97800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/16/2024
Initial Date FDA Received04/30/2024
Date Device Manufactured11/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
-
-