ZIMMER BIOMET, INC. ARCOS MODULAR REVISION SYSTEM 3.5MM HEX DRIVE; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2024
Event Type  malfunction  
Event Description
It was reported that the hex screwdriver broke while assembling the implant.There was no known impact or consequences to the patient.It was reported that no further information is available.
 
Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further information is available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the device had fractured at the tip.There are wear marks on the shaft in two places with one being near the fracture location.The features of the fracture surface visually align with zrm in which a bending overload fracture failure mode was identified.Complaint confirmed based on the evaluation of the returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS MODULAR REVISION SYSTEM 3.5MM HEX DRIVE
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19216903
MDR Text Key341431687
Report Number0001825034-2024-01109
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-301852
Device Lot NumberZB140501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2024
Initial Date FDA Received04/30/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-