Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/09/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the hex screwdriver broke while assembling the implant.There was no known impact or consequences to the patient.It was reported that no further information is available.
|
|
Manufacturer Narrative
|
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No further information is available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the device had fractured at the tip.There are wear marks on the shaft in two places with one being near the fracture location.The features of the fracture surface visually align with zrm in which a bending overload fracture failure mode was identified.Complaint confirmed based on the evaluation of the returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|