• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Failure to Charge (1085)
Patient Problem Cardiac Arrest (1762)
Event Date 08/17/2023
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device failed to charge.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Manufacturer Narrative
Zoll medical united kingdom evaluated the device and the device performed to specification.The device was put through extensive testing including functional testing, defib/pacer stress testing, bench handling, and defib cycling without duplicating the report.The device was recertified and returned to the customer.Review of the activity log showed the device detected a valid impedance.The device performed the first analysis and advised a shock and delivered the shock.The device conducted the next two analyses but did not recommend a shock and was advised to perform cpr.The user pressed the charge and shock buttons multiple times and the device neither charged the defib nor discharged the energy.The user powered the device on and off and the device was in aed mode.The user switched to manual mode to charge the device and deliver the energy.Review of the log indicated the user was unable to perform the charge operation due to the device being in aed mode.Per the design of the aed mode, the device is only able to charge at the selected energy (per the aed config) when the device detects a shockable rhythm and the zoll algorithm advises shock.The device does not charge energy by pressing the charge button in aed mode.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key19217279
MDR Text Key341437637
Report Number1220908-2024-01696
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/15/2024
Initial Date FDA Received04/30/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-