Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection, and functional tests of the returned device was performed following bwi procedures.Visual analysis revealed reddish material and a hole in the surface of the pebax.The temperature and impedance test were performed and the device was found working correctly.No temperature or impedance issues were observed.A patency test was performed, and the device was flushing correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal action were identified.The reddish material inside the pebax could be related to the temperature issue reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instruction for use states: when cleaning the tip electrode, be careful not to twist the tip electrode with respect to the catheter shaft; twisting may damage the tip electrode bond and loosen the tip electrode, or may damage the contact force sensor.In addition, in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle as part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref (b)(4).
|
It was reported that a patient underwent a paroxysmal atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter for which biosense webster¿s product analysis lab (pal) identified a hole on the pebax.It was initially reported by the customer that the temperature slope was too high on ngen generator.Blood was also observed under the surface of catheter tip which was not possible to wipe off.After replacing of the appropriate catheter cable the problem with this catheter was not solved.This catheter was replaced with another one and from the same type, so that procedure was ended successfully.There was no damage and consequences caused to the patient.There was no noise at all.The customer¿s reported temperature and blood inside the catheter are not issues that are considered to be mdr reportable since the potential that it could cause or contribute to a death or serious injury, or other significant adverse event, is remote.On 6-apr-2024, the bwi pal revealed that a visual inspection of the returned device found a hole in the pebax and a reddish material inside the pebax.These findings were reviewed and assessed the issue of a ¿hole¿ in the pebax as an mdr reportable malfunction since the integrity of the device has been compromised.
|