FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA Back to Search Results
Model Number OPT944
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo series humidifiers and are also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Optiflow + interfaces are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.Optiflow + interfaces are not intended for life support and appropriate patient monitoring must be used at all times.
 
Event Description
A healthcare facility in china reported that the opt944 optiflow + adult nasal cannula was heard to be leaking air between the tubing and the manifold.It was further reported by the healthcare facility that upon inspection, no visual signs of damage could be observed.There was no patient consequence.
 
Event Description
A healthcare facility in china reported that the opt944 optiflow + adult nasal cannula was heard to be leaking air between the tubing and the manifold.It was further reported by the healthcare facility that upon inspection, no visual signs of damage could be observed.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo series humidifiers and are also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Optiflow + interfaces are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.Optiflow + interfaces are not intended for life support and appropriate patient monitoring must be used at all times.Method: the complaint opt944 optiflow + adult nasal cannula was not returned to f&p for evaluation as the healthcare facility reported it was unavailable for return.Our investigation is based on the information and photograph provided by the healthcare facility, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the provided photograph revealed that the tubing of the opt944 optiflow + adult nasal cannula was slightly stretched near the manifold but did not confirm any damage of the tube that would result in a air leak.The healthcare facility reported that upon inspection of the subject device there was no detected damage that could cause a leak.Conclusion: without the return of the subject device and based on the information provided, our investigation was unable to determine any damage to the opt944 optiflow + adult nasal cannula that would result in an air leak.F&p's manufacturing controls for the optiflow + tubing include inspections during production for visual defects including cracks, tears, moulding defects, contamination, inclusions, discoloration and stretching or deformation.The subject opt944 optiflow + adult nasal cannula would have met the required specifications.The user instructions also state: "do not crush or stretch tube, to prevent loss of therapy." "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key19218516
MDR Text Key341589581
Report Number9611451-2024-00319
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012429117
UDI-Public(01)09420012429117(10)2102132567(11)220412
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot Number2102132567
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/01/2024
Initial Date FDA Received04/30/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-