This device is used for therapeutic purposes.(b)(4).Product evaluation: when received for analysis, there was evidence of biological contaminants [based off of the reactivity with hydrogen peroxide].When compared to the assembly drawing: the front hub showed a reduction in the outside diameter just proximal to the locking area; typically this damage has been associated with excess heat deformation.When viewed under magnification, there was some gouging of the inner shaft in the support area proximal to the tip.The shaft would turn by hand; however, it was not turning smoothly [it was gravely].The bur was installed into a handpiece, ran at 12,000 rpm in forward direction, and cut saw bone for 30 seconds; during the run the bur produced excess noise and vibration; however, it never stopped turning as reported.While removing the bur from the handpiece it was noticed that the front hub and outer tube radius were too hot to touch, even while wearing gloves.An attempt to measure the temperature using an infrared thermometer failed possibly due to the shape and size of the area being measured.The information indicates the shaft is binding.The alleged malfunction could not be confirmed, [bur tip got stuck / stopped turning] however; the excess heat may have resulted in the seizing of the bur over an extended period of time.Based off of the analysis findings alone, the underlying cause could not be determined.Method: simulated use testing.Microscopic inspection.
|