ARTHREX, INC. BIOCOMPOSITE INTERFERENC SCREW W/ DISP SHEATH; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number AR-1380C
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 12/13/2013
Event Type  Injury  
Event Description
The screw broke in half during acl procedure.Only half of the screw was placed, the other half was removed.Surgery was finished using a titanium screw to strengthen the structure and prevent another break.No further information available.
 
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested/is expected for evaluation but has not yet been received.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Complainant's event is typically caused by not inserting the implant co-axial to the bone tunnel, prying or leveraging while still loaded, improper bone preparation or flexing the joint during insertion.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If the device is returned and additional information is obtained, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOCOMPOSITE INTERFERENC SCREW W/ DISP SHEATH
Type of Device
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer Contact
vik bajnath, sr. mdr analyst.
1370 creekside boulevard
naples, FL 34108-1945
8009337017
MDR Report Key3561342
MDR Text Key4211165
Report Number1220246-2013-00282
Device Sequence Number1
Product Code MAI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K071176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/13/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/01/2015
Device Catalogue NumberAR-1380C
Device Lot Number600288
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/13/2013
Initial Date FDA Received01/08/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-