Model Number M00566731
Device Problems Occlusion Within Device (1423); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/19/2013
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an endovive safety peg kit push method was used during a percutaneous endoscopic gastrostomy procedure on (b)(6), 2013.According to the complainant, during the procedure, when the physician was tracking the peg tube over the guidewire, the guidewire would not advance through the peg tube.A second endovive safety peg kit push method was opened and the original guidewire was used to complete the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
Additional device info: the reported lot number could not be matched to the reported device; therefore, the lot expiration and device manufacture dates are unknown at this time.However, the complainant reported that the device was not expired.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE? SAFETY PEG KIT
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
ingrid matte
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key3570412
MDR Text Key4218399
Report Number3005099803-2014-00070
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberM00566731
Device Catalogue Number6673
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2013
Initial Date FDA Received01/13/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
-
-