Catalog Number AC020 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Code Available (3191)
|
Event Date 12/24/2013 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient had been implanted with crown sliders and skull pin on (b)(6) 2013.A surgery to remove the devices was performed on (b)(6) 3013, after an anterior skill pin had backed out.During the removal, the skull pin anterior head could not be disassembled from the slider, and so the devices were removed.There were no adverse patient consequences as a result, and the physician intends to perform another halo system installation procedure on (b)(6) 2013.
|
|
Manufacturer Narrative
|
A follow-up report will be submitted upon completion of the investigation.The incident in this report is associated with the device reported in mfr report # 1526439-2014-10546.
|
|
Manufacturer Narrative
|
Visual inspection noted that the two skull pins were still screwed into the 3d crown zero sliders.Various attempts to unscrew the skull pins were taken.However, it was very difficult to fully unscrew the skull pins.It appears that the skull pin threads have become ceased inside the sliders causing the skull pins not being able to unscrew.A review of the device history record (dhr) for the skull pin 2000 pkg-5-strl could not be conducted as the lot number etched on the skull pin does not match supplier ((b)(4)) records as it is an invalid lot number.Therefore, without the lot numbers, a review of the manufacturing records cannot be completed.A 12-month review of the complaint trend analysis for the skull pin 2000 pkg-5-strl was conducted on the specific product code from this complaint as the product family does not exist for this product.It was noted that there were no related complaints associated with the skull pin coming off and sliding; of which there was no harm to the patient but potential may exist if the fault were to recur.Review of complaints found no significant trends.The root cause cannot positively be determined.It is inconclusive of how the skull pin has loosened and slided off.No corrective action/preventive action is required as there has been no issue identified in the manufacturing or release of this device, and there have been no systematic trends.Therefore, this complaint will be closed with no further action required.
|
|
Search Alerts/Recalls
|
|
|