Catalog Number AC020
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Code Available (3191)
Event Date 12/24/2013
Event Type  malfunction  
Event Description
It was reported that a patient had been implanted with crown sliders and skull pin on (b)(6) 2013.A surgery to remove the devices was performed on (b)(6) 3013, after an anterior skill pin had backed out.During the removal, the skull pin anterior head could not be disassembled from the slider, and so the devices were removed.There were no adverse patient consequences as a result, and the physician intends to perform another halo system installation procedure on (b)(6) 2013.
 
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.The incident in this report is associated with the device reported in mfr report # 1526439-2014-10546.
 
Manufacturer Narrative
Visual inspection noted that the two skull pins were still screwed into the 3d crown zero sliders.Various attempts to unscrew the skull pins were taken.However, it was very difficult to fully unscrew the skull pins.It appears that the skull pin threads have become ceased inside the sliders causing the skull pins not being able to unscrew.A review of the device history record (dhr) for the skull pin 2000 pkg-5-strl could not be conducted as the lot number etched on the skull pin does not match supplier ((b)(4)) records as it is an invalid lot number.Therefore, without the lot numbers, a review of the manufacturing records cannot be completed.A 12-month review of the complaint trend analysis for the skull pin 2000 pkg-5-strl was conducted on the specific product code from this complaint as the product family does not exist for this product.It was noted that there were no related complaints associated with the skull pin coming off and sliding; of which there was no harm to the patient but potential may exist if the fault were to recur.Review of complaints found no significant trends.The root cause cannot positively be determined.It is inconclusive of how the skull pin has loosened and slided off.No corrective action/preventive action is required as there has been no issue identified in the manufacturing or release of this device, and there have been no systematic trends.Therefore, this complaint will be closed with no further action required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SKULL PIN 2000 PKG-5-STRL
Type of Device
COMPONENT, TRACTION, INVASIVE
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
Manufacturer Contact
anita barnick
325 paramount drive
raynham, MA 02767
5088283583
MDR Report Key3575961
MDR Text Key17160198
Report Number1526439-2014-10544
Device Sequence Number1
Product Code JEC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK993099
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/24/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAC020
Device Lot Number16488
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2014
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date12/24/2013
Event Location Hospital
Initial Date Manufacturer Received 03/11/2014
Initial Date FDA Received01/15/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/25/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Patient Sequence Number1
Treatment
3D CROWN ZERO DEG SLIDER
-
-