Discordant, falsely low chloride results were obtained on eight patient samples on an advia 1800 instrument.The discordant chloride results were reported to the physician(s).The customer repeated the samples and those results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low chloride results.
|
A siemens healthcare diagnostics cse (customer service engineer) was dispatched to the customer site for instrument evaluation.After evaluating the instrument and data, the cse determined that the cause of the discordant chloride results was unknown.The cse removed and cleaned the ion selective electrode module, replaced the chloride electrode, performed calibration and ran qc.The instrument is performing within specifications.Further evaluation of the device is not required.
|