Complainant alleged that during patient use, the autopulse® resuscitation system would not provide continuous compressions.The autopulse performed one compression, stopped and displayed a "realign patient" message.Customer attempted to realign the patient to the designated area on the platform, however this did not resolve the issue.No adverse patient sequelae was reported.No further details were provided.
|
Customer also reported that after the incident, the autopulse was tested with a mannequin and the same issue occurred.The autopulse platform (b)(4) was returned for evaluation.The reported complaint of the platform stopping compressions and displaying a message stating "realign patient" could not be confirmed.The platform was run for 20 minutes with a test mannequin and an additional 7 minutes with a lrtf (large resuscitation test fixture) and no faults were found.A review of the archive shows that on (b)(6) 2013, numerous user advisory 2 faults (compression tracking error) had occurred and that a small patient or object was used on the platform.Per the autopulse® resuscitation system model 100 user guide (pn: 12555-001), "the autopulse enters a user advisory state or the fault state when one of several conditions is detected.A user advisory generally indicates that a misalignment or inappropriate movement of the patient or the lifeband has occurred.A fault generally indicates that the autopulse has detected an inappropriate internal condition.Both conditions are typically correctable by the operator.Follow the instructions on the screen and then attempt to restart active operation by pressing the start/continue button." as this fault occurred approximately one month prior to the reported event date of (b)(6) 2013, it is deemed unrelated to the reported complaint.
|