It was reported there was a problem with a dual lumen cannula.The guidewire became knotted at the end of the wire after it was inside the patient and the surgeon could not remove the wire.The surgeon was able to remove the wire with the introducer, un-knot the wire and then place it back in to remove the wire.The customer felt the wire was too flimsy.It was later reported that the patient expired due to heart failure because a transplant recipient could not be found in time.(b)(4).Ref# mfr 8010762-2014-00030.
|