Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Information (3190)
Event Date 01/10/2014
Event Type  Injury  
Event Description
It was reported patient underwent an initial procedure on (b)(6) 2014.During the procedure, the tip of the speedpass lariat fractured inside the patient.The fractured tip was removed from the patient.
 
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.Device availability - the device is reported to be available for evaluation; however, it has not been received by biomet orthopedics to date.In the event that the device is received and evaluated, a follow up report will be sent to the fda to provide results.
 
Manufacturer Narrative
Examination of returned device found no evidence of product non-conformance.Dimensional evaluation found component to be within appropriate design specification.The root cause of the event could not be determined due to insufficient information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPDPASS LARIAT W/NIT 25DEG RT
Type of Device
PASSER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
amanda zajicek
56 e. bell drive
warsaw, IN 46582
5743726782
MDR Report Key3612176
MDR Text Key4069726
Report Number0001825034-2014-00729
Device Sequence Number1
Product Code HWQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date11/30/2014
Device Model NumberN/A
Device Catalogue Number904012
Device Lot Number523840
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2014
Initial Date FDA Received02/07/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/23/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-