Model Number 20
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 01/13/2014
Event Type  malfunction  
Event Description
It was reported that the therapy connector assembly on the customer's device had a broken connector pin stuck inside from the therapy cable assembly.A broken pin stuck within the therapy connector would prohibit the connection of another therapy cable assembly and would potentially prohibit defibrillation therapy delivery from the device.There was no patient use associated with the reported failure.
 
Manufacturer Narrative
(b)(4).Physio-control evaluated the device and verified the reported failure.Physio replaced the therapy connector assembly and observed proper device operation through functional and performance testing.The device was then returned to the customer for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key3612649
MDR Text Key4172883
Report Number3015876-2014-00104
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2014
Device Age9 YR
Event Location Hospital
Initial Date Manufacturer Received 01/13/2014
Initial Date FDA Received02/07/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-