Model Number ICM120V4
Device Problem Split (2537)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2013
Event Type  malfunction  
Event Description
The reporter indicated the surgeon attempted to insert a 12.0mm icm120v4 implantable collamer lens in the patient's left eye (os) but the surgeon noted a split on the lens and the lens was not inserted.The lens touched the patient's eye, there was no patient injury.The backup lens was implanted.
 
Manufacturer Narrative
(b)(4).Lens not returned.(b)(4).
 
Manufacturer Narrative
Method: work order search.Results: a lens work order search was performed and no similar complaints were found within the work order.Visual inspection of the returned product found no visible damage to the lens.The lens was returned liquid.The lens was rehydrated in bss for re-measurement.The lens length was measured and the result of the measurement was compared against the original value and the lens was found to be in specification.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker ave
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker ave
monrovia CA 91016
Manufacturer Contact
althea watson
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key3625684
MDR Text Key4134357
Report Number2023826-2014-00108
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/30/2016
Device Model NumberICM120V4
Other Device ID NumberDIOPTER -10.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/16/2014
Initial Date FDA Received02/12/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/16/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INJECTOR MODEL AND LOT NUMBER UNK; CARTRIDGE MODEL AND LOT NUMBER UNK
Patient Age33 YR
-
-